Chamber FAQ

GR-326-CORE for fiber optic connectors

You will need an insertion / return loss test system to go along with our chamber to do these tests.  We recommend the Optical Component Environmental Test System from JDS Uniphase. It is ready to work with an ESPEC environmental chamber, making the perfect optical components environmental test system for long term testing to Telcordia (Bellcore) standards GR-326, 1209, 1221. 

 

All the required tests can be done by our Platinous model

 

                                                                                                               
ParagraphTitleDescriptionBenchtop ModelsReach-in Models
4.4.2.1Thermal Age Test85°CAny benchtop or PH or PV ovensAny Platinous or Platinum
4.4.2.2Thermal Cycle Test -40 to 75°C Any Platinous or Platinum
4.4.2.3Humidity Aging Test75°C/95%Any benchtop or PH or PV ovensAny Platinous or Platinum  
4.4.2.4Humidity/Condensation Cycling Test -10 to 65°C w/ condensation see note belowAny Platinous or Platinum
4.4.2.5Dry-out Step75°CAny benchtop or PH or PV ovensAny Platinous or Platinum
4.4.4.1Dust TestArizona fine dust  
4.4.4.2Adhesive Test65°CAny benchtop or PH or PV ovensAny Platinous or Platinum

 

For test 4.4.2.4, we don’t recommend our benchtop series due to the requirement for heat-up as fast as possible (3.75°C/min. minimum) in order to create condensation.  For Platinous models, a special, but inexpensive modification is required.  